Serkan Aslan1, Aysel Türkvatan2, Ahmet Güner1, Serkan Kahraman1, Ümit Bulut1, Gökhan Demirci1, Enes Arslan1, Ömer Çelik1, Mehmet Ertürk1

1Clinic of Cardiology, University of Health Sciences İstanbul Mehmet Akif Ersoy Thoracic and Cardiovascular Surgery Training and Research Hospital, İstanbul, Türkiye
2Clinic of Radiology, University of Health Sciences İstanbul Mehmet Akif Ersoy Thoracic and Cardiovascular Surgery Training and Research Hospital, İstanbul, Türkiye

Keywords: Aortic angulation; computed tomography; self-expanding valve; transcatheter aortic valve implantation

Abstract

Introduction: Aortic angulation (AA), defined as the angle between the aortic annulus plane and the horizontal plane, may result in failed prosthesis positioning. The effect of AA on the procedural and short-term outcomes with the portico valves for transcatheter aortic valve implantation (TAVI) has not been fully investigated. The present study aimed to evaluate the impact of AA on device success and early outcomes of TAVI using a selfexpanding portico valve.

Patients and Methods: Preoperative computed tomography scans of 121 consecutive patients treated with the portico valve were analyzed. TAVI device success and outcomes were determined according to VARC-3 definitions. Patients were divided into two groups based on mean AA.

Results: The mean AA was 48.7 ± 8.9°. There were no differences in technical success (92.2 vs 89.5%, p= 0.604), device success (81.2% vs 77.2%, p= 0.582), and early safety endpoints (68.8% vs 61.4%, p= 0.397) between the AA≤ 48° and AA>48° groups. The frequency and severity of paravalvular aortic regurgitation (PAR) was statistically higher in patients with AA> 48° (p= 0.028). Moreover, an increased AA was also associated with valve malposition (12.3% vs. 1.6%, p= 0.018), prolonged procedure time (85 ± 26 vs. 75 ± 20 minutes, p= 0.028), and greater Δimplantation depth (2.2 ± 0.5 vs. 0.8 ± 0.1 mm; p<0.001).

Conclusion: Despite comparable device success and early outcomes rates, increased AA is associated with higher rates of PAR and valve malposition, with the self-expanding portico valve.

Introduction

Transcatheter aortic valve implantation (TAVI) has developed rapidly in the last two decades as an alternative treatment option for patients with severe aortic stenosis who are at intermediate-to-high or prohibitive risk for surgery, and it is expected to expand in the future to the treatment lowerrisk patients(1-4). Despite the role of advances in implantation techniques and/or prosthesis design in reducing TAVI complications and the promising results of clinical trials, particular anatomical circumstances may pose challenges for the proper positioning and consequent functioning of self-expandable TAVI valves(5,6).

Aortic angulation (AA), quantified by computed tomography angiography (CTA) and defined as the angle between the aortic annulus plane and the horizontal plane (Figure 1), can make successful positioning and optimal placement of the selfexpanding prosthesis more challenging, especially in cases of the excessively angled aortic root(7,8). Several studies using different valve designs have produced contradictory results regarding the effect of increased AA on TAVI outcomes(9-11). Nevertheless, data on predictors of device success of the Portico transcatheter heart valve (THV) system (Abbott Vascular, Santa Clara, CA, USA) after TAVI is scarce; particularly, the effect of increased AA on the acute procedural and early outcomes has not been fully investigated.

The present study aimed to evaluate the impact of AA on device success, and procedural and short-term outcomes in patients who underwent TAVI with a self-expanding portico valve.

Materials and Methods

Study Population and Design

We conducted a retrospective review of 128 consecutive patients who underwent a TAVI with a self-expanding Portico TAVI system at our hospital from March 2017 to October 2021. A total of seven patients were excluded from the analysis for the following reasons: Bicuspid valve (n= 5), previously implanted surgical bioprosthetic valves (n= 1), and inability to access all medical records (n= 1). The remaining 121 patients represented the final study population. Consistent with the previously published research, the study population was divided into two groups based on the mean AA to evaluate whether AA affected clinical outcomes in patients undergoing TAVI. All baseline demographic and clinical information, echocardiography and CTA imaging data, procedural details, and 30-day outcomes were obtained from the institutional TAVI database and compared between the two groups. This study was approved by the institutional review board and complied with the Declaration of Helsinki.

TAVI Procedure

All patients were discussed in our institutional Heart Team meetings and were considered at high risk for valve surgery. All TAVI procedures were performed in a hybrid catheterization laboratory under local anesthesia with conscious sedation and were performed by experienced, formally trained physicians. A pre-shaped extra stiff Safari Guidewire (Boston Scientific, Marlborough, Massachusetts, USA) was used in all patients. The portico valve was implanted under fluoroscopic guidance via transfemoral access as previously described(12). Post-dilatation under rapid pacing was considered in the case of incomplete frame expansion or remaining moderate or severe paravalvular aortic regurgitation (PAR). A final control was performed by aortography. Implantation depth was assessed in the implantation projection by fluoroscopy as the length of the stent frame from the aortic annular plane to the left ventricular outflow tract (LVOT), measured at the level of the noncoronary cusp (NCC) and left coronary cusp (LCC). ΔImplantation depth was calculated as the length of the LCC implantation depth minus the NCC implantation depth.

CTA Data Acquisition and Image Analysis

CTA acquisitions of all patients were performed using a 320-row CTA scanner (Aquilion ONE, Toshiba Medical Systems, Otawara, Japan). All CTA examinations were evaluated using a dedicated workstation (Vitrea version 6.4, Vital Images, USA) to allow for multiplanar reconstruction analysis. Since aortic root dimensions are typically larger in systole, a double oblique transverse view (en face view of the aortic valve) was created during the systolic phase (20-40%) of the cardiac cycle to measure the aortic root dimensions. The aortic annulus was defined as the virtual basal ring aligned with the most basal attachments (basal hinge points) of the three aortic valve cusps on a double-oblique transverse view. First, the minimum and maximum diameters of the annulus were obtained. The annular perimeter was manually drawn using a planimetry tool, after which the area and circumference of the annulus were derived by a workstation software. The percentage of oversizing was calculated with the following equation: (prosthesis/annulus perimeter - 1)*100(13). The eccentricity index was calculated with the following formula: 1 - (minimum diameter/maximum diameter)(14). The degree of aortic valve commissural calcification was visually graded from 0 to 3 (0= no calcification, 1= single lesions< 5 mm, 2= calcium lesions> 5 mm or affecting 2 or more leaflets, 3= severe calcification affecting all 3 leaflets) by a semi-quantitative model as described in the previous publication(15). AA was measured from the implantation projection derived from the reconstruction of CTA and was defined as the angle between the plane of the aortic annulus and the horizontal plane.

Echocardiographic Study

The experienced physicians who were blinded to the patients’ clinical characteristics performed all the echocardiographic measurements at baseline and within 30 days. The severity of aortic stenosis, left ventricular ejection fraction, or PAR were evaluated according to the guidelines(16). The fiveclass grading scheme (mild, mild-moderate, moderate, moderate-severe, and severe) recommended by Valve Academic Research Consortium (VARC)-3 was used to define residual PAR(17). For the purposes of this paper, trivial jets were classified as having no PAR, whereas moderate and severe PAR were grouped together.

Definition of Endpoints of the Study

TAVI device success, the intra-, and postoperative data, and 30-day clinical outcomes were defined according to the standardized criteria proposed by VARC-3(17). The primary endpoint of the present study was device success, which was defined as the combined endpoint of technical success, freedom from allcause mortality (within 30 days), absence of surgery or intervention related to the procedural complications, and intended valve performance (mean gradient< 20 mmHg, and less than moderate PAR). The secondary endpoints of the study included clinical outcomes of all-cause mortality, stroke, acute kidney injury, myocardial infarction, major vascular complications, life-threatening bleeding, moderate or severe PAR, and new pacemaker implantation within 30-days or during the index hospitalization stay.

Statistical Analysis

The normality of the sample data distribution was tested visually (with histograms and probability curves) or statistically (with Kolmogorov-Smirnov and Shapiro-Wilk tests). Continuous variables were summarized using mean ± standard deviation or median [interquartile range (IQR)] and were compared by using the Student’s t-test or Mann-Whitney U test, where appropriate. Categorical and dichotomous variables are presented as frequencies and percentages and were compared by using Pearson Chi-square or Fisher exact test, as appropriate. Receiver operating characteristic (ROC) analyses were performed to determine the cut-off value of AA that provided the best combination of sensitivity and specificity for device success. All statistical analyses were two-tailed tests with p< 0.05 considered statistically significant. The statistical analyses were conducted with the SPSS software, version 24.0 (SPSS Inc., Chicago, Illinois, United States).

Results

The baseline clinical patient characteristics, echocardiographic, and pre-procedural CTA imaging details of the study population are displayed in Table 1. The mean age of the study population was 80.2 ± 6.2 years; 36.4% were male and had a median STS score of 6.2 (4.0-8.9). Coronary artery disease (71.1%) was the most common comorbidity in TAVI patients, followed by hypertension (64.5%) and diabetes (42.1%). The mean AA was 48.7 ± 8.9°. This value was used as a cut-off to identify the two study groups (AA≤ 48°: 64 and AA> 48°: 57 patients). Baseline patient and pre-TAVI imaging characteristics were comparable between both groups. The analysis is detailed in Table 1.

Procedural details and acute perioperative outcomes are displayed in Table 2. Δimplantation depth was found to be statistically higher in patients with AA> 48° compared with those with AA≤ 48° (2.2 ± 0.5 mm vs 0.8 ± 0.1 mm; p< 0.001). The procedure time was significantly (10 min) longer in the higher AA group (85 ± 26 minutes in its favor when compared to the lower AA group 75 ± 20 minutes, p= 0.028). The prolonged procedure time in the higher AA group was due to the extra time required to implant the second valve in the event of valve malposition or moderate-to-severe PAR. In eight of 121 (6.6%) patients, the implant was initially unsuccessful. Malposition of the prosthesis was significantly higher among patients with increased AA (>48°: seven of 57 patients, 12.3% vs. ≤48°: one of 64 patients, 1.6%, p= 0.018). In four of these cases in the higher AA group, the valve prosthesis moved up upward, within the aortic annulus following initial correct positioning. In the other three patients, the valve prosthesis moved upward into the ascending aorta after complete deployment and could not be retrieved. The valves were withdrawn using a snare technique and were fixed within the ascending aorta, without affecting the coronaries. Only one patient in the ≤48° group developed valve migration requiring a second valve. The need for the second valve was numerically higher in patients with AA> 48° compared with those with AA≤ 48° (seven of 57 patients, 12.3% vs. three of 64 patients, 4.7%, p= 0.188), but did not reach significance due to case numbers. In summary, technical success was achieved in 90.9% of patients and was comparable between the two groups (AA> 48°: 89.5% vs. AA≤ 48°: 92.2%, p= 0.604) (Table 3). In the overall cohort, device success at day 30 was 79.3%. No significant difference was found between groups for the device success, the primary endpoint of the study (AA> 48°: 77.2% vs. AA≤ 48°: 81.2%, p= 0.582). As shown in Figure 2, the AUC of 0.514 for AA of >50° to predict device success failed to distinguish between those with and without device success (AUC= 0.514; 95% CI= 0.385-0.642, p= 0.835). There was no significant difference in early safety endpoints between AA groups (AA> 48°: 61.4% vs. AA≤ 48°: 68.8%, p= 0.397). Further details are provided in Table 3.



One of the primary findings of this study was that the frequency and severity of PAR, as measured by the VARC-3, were statistically higher in the higher AA group (p= 0.028). Figure 3 illustrates the frequency and severity of PAR. Of the 118 patients with interpretable echocardiography studies within 30 days, PAR grading was determined in 76 patients (64.4%), with no subject presenting with severe PAR. In the higher AA group, PAR was none/trace in 22.2%, mild in 44.4%, mild-moderate in 25.9%, and moderate in 7.4%. In the lower AA group, PAR was none/ trace in 46.9%, mild in 32.8%, and mild-moderate in 18.8%. Only one (1.6%) patient had moderate PAR. These findings suggest that there is a relationship between the severity of PAR and AA.

Discussion

The principal findings of this investigation are as follows:

1) There is no statistically significant relationship between increased AA and TAVI device success rates or short-term outcomes,

2) Increased AA is associated with more frequency and severity of PAR, higher rates of valve malposition, and prolonged procedure time,

3) There is a statistically significant relationship between AA and Δimplantation depth. To the best of our knowledge, this is the first study focusing on the Portico valve system performance based on AA, with a cut-off value of >48° consistent with previous reports.

Increased AA causes the valve to be subjected to a higher degree of bending, which can make it difficult to position the valve properly. Asymmetric positioning of the valve may affect the success of the acute procedure adversely. The fact that the self-expanding valves have a longer stent frame and the delivery systems lack active flexion and extension makes it more difficult to optimize the co-axial alignment of these prostheses than balloon-expandable valves. It has been clarified that the high-angle aorta negatively affects TAVI results with old-generation self-expanding prostheses(7,8,18). Previously, Abramowitz et al. reported in their study evaluating the effect of increased AA on TAVI outcomes that increased AA adversely affected acute procedural success after the self-expandable valve, nevertheless, it did not affect TAVI results after affixing the balloon-expandable valve(7). On the other hand, several studies conducted in collaboration with CoreValve reported conflicting results regarding the effect of AA on device success rates between higher and lower groups(9,10,19). A subsequent study conducted by Popma et al. on self-expandable valves reported that there is no association between AA and procedural success or clinical outcomes. The authors attributed these findings to the use of the most up-todate techniques in valve placement(19). The differences in these results can probably be explained by the technical improvement in new-generation devices.

Only one study has examined the effect of aortic root opening on TAVI outcomes in patients who have been implanted with the portico valve. Gorla et al. evaluated this in a retrospective study of 392 patients who had undergone TAVI with Portico, Evolut-R, and Acurate-NEO(20). Postoperative device success was specified based on the VARC-2 definition and was generally satisfactory for the Portico valve and was comparable in both the horizontal (95.0%) and non-horizontal (86.8%) groups. Similar to the previous study, primary device success in our study was not affected by AA and was comparable in both groups (AA≤ 48°: 81.2% vs. AA> 48°: 77.2%; p= 0.582). The results are relatively variable due to differences in the definition of device success in different studies. The reason why our device success proved to be lower than the other study is that we judged it according to the definition of device success criteria, which was updated in 2021 with the VARC-3 definitions. On the other hand, all other composite endpoints, including technical success and early safety were similar between both AA groups in our study.

The study conducted by Gorla et al. also examined the relationship between PAR and AA after TAVI(20). It reported that in portico valves there is no association between AA and the incidence of moderate-to-severe PAR. However, the authors stated that these findings may have been affected by the low number of patients in the portico group (20 patients). Indeed, there were no patients who developed moderate to severe PAR after TAVI in the portico horizontal group in their study. The results of the study mentioned above regarding the relationship between AA angle and PAR were not confirmed by our findings. We identified a significant correlation between high AA and increased frequency and severity of PAR. In our study, we used the five-class grading scheme recommended by VARC-3, which can be easily narrowed down to the three-class scheme when needed in order to reduce variability between echocardiography laboratories(17,21). The moderate PAR rate in our study was 4.2%, which is in line with previous reports(22-25). With only five cases with moderate PAR after TAVI, no multivariate analysis regarding an AA and other variables as independent predictors for moderate to severe PAR could be offered. In our study, we hypothesize that asymmetrical positioning of the prosthetic stent frame within the implantation site and a consequent greater Δimplantation depth is also associated with the risk of PAR greater than mild in patients with TAVI. Briefly, the implantation depth was measured to be less on the noncoronary sinus side and more on the left coronary sinus side in patients with high AA. It was found to be statistically higher in patients with high implantation depth, which is an indication of asymmetrical localization (2.2 ± 0.5 mm vs. 0.8 ± 0.1 mm; p< 0.001).

Previously, Schultz et al. hypothesized that the apposition force exerted by the long prosthesis frame on the adjacent tissue of the annulus would be higher on the outer curve of the aorta or the right side of the annulus, and lower on the inner curve of the aorta or the left side of the annulus(26). They supported their hypothesis by detecting a higher malposition rate at the implantation site within the curve. Increased AA may affect the radial strength of the prosthesis and its ability to completely seal the paravalvular space during TAVI. This increases the possibility of post-implantation complications such as paravalvular leakage, valve embolization, and THV-in-THV. Indeed, several procedural success criteria were found to be significantly different in the high AA group: Paravalvular leak, increased valve malposition rates, and prolonged procedure time. The results of our study confirmed the negative effect of increased AA after TAVI on prosthesis malposition, possibly resulting from the decrease in grip strength due to the insufficient apposition force of the frame and asymmetrical placement.

In fact, the design of the new generation of prostheses aims to simplify and standardize the TAVI procedure even in the most challenging anatomical scenarios. The self-expanding Portico THV system has recently undergone several improvements to the carrier unit (FlexNav, Abbott Cardiovascular) to support especially more precise valve placement and increase the procedural safety and accuracy of the platform. In a recent study by Fontana et al. Portico valve implantation with FlexNav was associated with an excellent safety profile and a technical device success of 96.7% within 30 days. Echocardiography revealed a moderate PAR of 4.1% on the 30th day(27). However, 3.3% of subjects required a second valve to be implanted during the procedure. The overall success and all these added features in patients undergoing Portico transcatheter heart valve implantation with the new FlexNav delivery system now need to be tested in more specific and challenging anatomical scenarios such as in the presence of an angled aorta.

It is important to emphasize limitations pertinent to the methods of this study. First of all, this was a single-center retrospective study and included a relatively small patient population. However, it should be noted that it is currently the highest reported single-center study with the Portico valve in this field. Second, due to the limited number of patients with moderate/severe PAR, we were unable to perform a multivariate regression analysis on statistically significant univariate factors to determine whether AA is an independent predictor of PAR. Lastly, only the shortterm follow-up data have been included. The present study focuses on device success and early composite endpoints. Future prospective, randomized studies with a larger number of patients and longer follow-up may clarify this subject.

In conclusion, the current study suggests that AA does not significantly affect device success or early clinical outcomes following TAVI with the self-expanding Portico THV system. Increased AA is associated with more frequency and severity of PAR, higher rates of valve malposition, and prolonged procedure time. Portico with the FlexNav delivery system, which facilitates control of valve delivery and predictable valve placement, may result in improved outcomes. Nevertheless, further large-scale studies are warranted to validate the findings of this study.

Cite this article as: Aslan S, Türkvatan A, Güner A, Kahraman S, Bulut Ü, Demirci G, et al. Effect of aortic angulation on outcomes in transcatheter aortic valve implantation with the self-expanding portico valve. Koşuyolu Heart J 2022;25(2):132-140.

Ethics Committee Approval

The approval for this study was obtained from University of Health Sciences İstanbul Mehmet Akif Ersoy Thoracic and Cardiovascular Surgery Training and Research Hospital Ethics Committee (Decision no: 2021-73, Date: 28.09.2021).

Peer Review

Externally peer-reviewed.

Author Contributions

Concept/Design - SA, ME; Analysis/Interpretation - SA, AT, ÜB; Data Collection - SA, ÜB, GD, EA; Writing - SA, AT, AG; Critical Revision - AG, SK, ÖÇ; Final Approval - ME; Statistical Analysis - SA, SK; Overall Responsibility - SA.

Conflict of Interest

The authors declared that there was no conflict of interest during the preparation and publication of this article.

Financial Disclosure

The authors declared that this study has received no financial support.

References

  1. Smith CR, Leon MB, Mack MJ, Miller DC, Moses JW, Svensson LG, et al. Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med 2011;364:2187-98. [Crossref]
  2. Thourani VH, Kodali S, Makkar RR, Herrmann HC, Williams M, Babaliaros V, et al. Transcatheter aortic valve replacement versus surgical valve replacement in intermediate-risk patients: A propensity score analysis. Lancet 2016;387:2218-25. [Crossref]
  3. Mack MJ, Leon MB, Thourani VH, Makkar R, Kodali SK, Russo M, et al. Transcatheter aortic-valve replacement with a balloon-expandable valve in low-risk patients. N Engl J Med 2019;380:1695-705. [Crossref]
  4. Popma JJ, Deeb GM, Yakubov SJ, Mumtaz M, Gada H, O’Hair D, et al. Transcatheter aortic-valve replacement with a self-expanding valve in low-risk patients. N Engl J Med 2019;380:1706-15. [Crossref]
  5. Van Mieghem NM, Popma JJ, Deeb GM, Yakubov SJ, Serruys PW, Windecker S, et al. Complete 2-year results confirm bayesian analysis of the SURTAVI trial. JACC Cardiovasc Interv 2020;13:323-31. [Crossref]
  6. Carroll JD, Mack MJ, Vemulapalli S, Herrmann HC, Gleason TG, Hanzel G, et al. STS-ACC TVT registry of transcatheter aortic valve replacement. Ann Thorac Surg 2021;111:701-22. [Crossref]
  7. Abramowitz Y, Maeno Y, Chakravarty T, Kazuno Y, Takahashi N, Kawamori H, et al. Aortic angulation attenuates procedural success following self-expandable but not balloon-expandable TAVR. JACC Cardiovasc Imaging 2016;9:964-72. [Crossref]
  8. Veulemans V, Maier O, Bosbach G, Polzin A, Piayda K, Afzal S, et al. Novel insights on outcome in horizontal aorta with self-expandable newgeneration transcatheter aortic valve replacement devices. Catheter Cardiovasc Interv 2020;96:1511-19. [Crossref]
  9. Bob-Manuel T, Pour-Ghaz I, Sharma A, Chinta VR, Abader P, Paulus B, et al. Correlation between aortic angulation and outcomes of transcatheter aortic valve replacement with new-generation valves. Curr Probl Cardiol 2021;46:100415. [Crossref]
  10. Rashid HN, McCormick LM, Talman AH, Ihdayhid AR, Nerlekar N, Amiruddin AS, et al. Effect of aorto-ventricular angulation on procedural success in transcatheter aortic valve replacements with the lotus valve system. Catheter Cardiovasc Interv 2018;91:1365-70. [Crossref]
  11. Medranda GA, Musallam A, Zhang C, Rappaport H, Gallino PE, Case BC, et al. The impact of aortic angulation on contemporary transcatheter aortic valve replacement outcomes. JACC Cardiovasc Interv 2021;14:1209-15. [Crossref]
  12. Tzikas A, Amrane H, Bedogni F, Brambilla N, Kefer J, Manoharan G, et al. Transcatheter aortic valve replacement using the Portico system: 10 things to remember. J Interv Cardiol 2016;29:523-9. [Crossref]
  13. Blanke P, Pibarot P, Hahn R, Weissman N, Kodali S, Thourani V, et al. Computed tomography-based oversizing degrees and incidence of paravalvular regurgitation of a new generation transcatheter heart valve. JACC Cardiovasc Interv 2017;10:810-20. [Crossref]
  14. Wong DT, Bertaso AG, Liew GY, Thomson VS, Cunnington MS, Richardson JD, et al. Relationship of aortic annular eccentricity and paravalvular regurgitation post transcatheter aortic valve implantation with CoreValve. J Invasive Cardiol 2013;25:190-5.
  15. Marwan M, Achenbach S, Ensminger SM, Pflederer T, Ropers D, Ludwig J, et al. CT predictors of post-procedural aortic regurgitation in patients referred for transcatheter aortic valve implantation: An analysis of 105 patients. Int J Cardiovasc Imaging 2013;29:1191-8. [Crossref]
  16. Lang RM, Badano LP, Mor-Avi V, Afilalo J, Armstrong A, Ernande L, et al. Recommendations for cardiac chamber quantification by echocardiography in adults: An update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J Am Soc Echocardiogr 2015;28:1-39.e14. [Crossref]
  17. Généreux P, Piazza N, Alu MC, Nazif T, Hahn RT, Pibarot P, et al. Valve academic research consortium 3: Updated endpoint definitions for aortic valve clinical research. Eur Heart J 2021;42:1825-57. [Crossref]
  18. Roule V, Placente A, Sabatier R, Bignon M, Saplacan V, Ivascau C, et al. Angles between the aortic root and the left ventricle assessed by MDCT are associated with the risk of aortic regurgitation after transcatheter aortic valve replacement. Heart Vessels 2018;33:58-65. [Crossref]
  19. Popma JJ, Reardon MJ, Yakubov SJ, Hermiller JB Jr, Harrison JK, Gleason TG, et al. Safety and efficacy of self-expanding tavr in patients with aortoventricular angulation. JACC Cardiovasc Imaging 2016;9:973-81. [Crossref]
  20. Gorla R, De Marco F, Garatti A, Bianchi G, Popolo Rubbio A, Acerbi E, et al. Impact of aortic angle on transcatheter aortic valve implantation outcome with Evolut-R, Portico, and Acurate-NEO. Catheter Cardiovasc Interv 2021;97:135-45. [Crossref]
  21. Hahn RT, Pibarot P, Weissman NJ, Rodriguez L, Jaber WA. Assessment of paravalvular aortic regurgitation after transcatheter aortic valve replacement: Intra-core laboratory variability. J Am Soc Echocardiogr 2015;28:415-22. [Crossref]
  22. Maisano F, Worthley S, Rodés-Cabau J, Linke AH, Fichtlscherer S, Schäfer U, et al. Early commercial experience from transcatheter aortic valve implantation using the Portico ™ bioprosthetic valve: 30- day outcomes in the multicentre PORTİCO-1 study. EuroIntervention 2018;14:886-93. [Crossref]
  23. Linke A, Holzhey D, Möllmann H, Manoharan G, Schäfer U, Frerker C, et al. Treatment of aortic stenosis with a self-expanding, resheathable transcatheter valve: One-year results of the international multicenter Portico transcatheter aortic valve implantation system study. Circ Cardiovasc Interv 2018;11:e005206. [Crossref]
  24. Möllmann H, Linke A, Holzhey DM, Walther T, Manoharan G, Schäfer U, et al. Implantation and 30-day follow-up on all 4 valve sizes within the Portico transcatheter aortic bioprosthetic family. JACC Cardiovasc Interv 2017;10:1538-47. [Crossref]
  25. Manoharan G, Linke A, Moellmann H, Redwood S, Frerker C, Kovac J, et al. Multicentre clinical study evaluating a novel resheathable annular functioning self-expanding transcatheter aortic valve system: Safety and performance results at 30 days with the Portico system. EuroIntervention 2016;12:768-74. [Crossref]
  26. Schultz CJ, Tzikas A, Moelker A, Rossi A, Nuis RJ, Geleijnse MM, et al. Correlates on MSCT of paravalvular aortic regurgitation after transcatheter aortic valve implantation using the Medtronic CoreValve prosthesis. Catheter Cardiovasc Interv 2011;78:446-55. [Crossref]
  27. Fontana GP, Bedogni F, Groh M, Smith D, Chehab BM, Garrett HE Jr, et al. Safety profile of an intra-annular self-expanding transcatheter aortic valve and next-generation low-profile delivery system. JACC Cardiovasc Interv 2020;13:2467-78. [Crossref]

Figure and Tables